CONTOUR® NEXT ONE
Glucose Meter
PAGE 38
REMOTE CARE PARTNERS
V1.0
TECHNICAL INFORMATION: PRECISION
A measurement repeatability study was
conducted with the Contour Next ONE blood
glucose monitoring system using 5 venous whole
blood specimens with glucose levels from 43
mg/dL to 333 mg/dL. With each blood specimen,
each of 3 lots of Contour Next test strips was
tested 10 times on each of 10 instruments for a
total of 300 readings. The following precision
results were obtained.
Table 1: System repeatability results for Contour Next ONE
meter using Contour Next test strips