3.2. Draft measurement
The draft present in the flue gas channel has already been indicated
by the trend display earlier. The chimney draft value is not stored by
pressing
<m>
as the differential pressure sensor is very sensitive and
drifts. For this reason it is recommended to re-calibrat the sensor
immediately before a measurement to ensure a good accuracy.
A differential pressure measurement is started by pressing
<DRAFT>
.
The current value is displayed as well as a reminder to reset zero
point. Remove the draft tubing and press the
<MODE>
key. The sen-
sor will now be re-calibrated.
Fix the tubing. The display now shows the exact value which can
saved value
:
-0.08 hPa
be stored by pressing
<m>
. This value will be added to those already
stored. The value saved is displayed. To leave the draft measurement
press
<ESC>
.
Draft measurement
Draft
- 0,05hPa
Reset zero point : release draft
hose, then press <MODE> !