Programming
CM
Rev. B 06/2011
30
<CR> <LF> (0D
h
0A
h
) is closing the answer.
Error message:
*Syntax Error “*“ is the character for “Error“
.
8.2
Device Setup
8.2.1
Temperature Calculation
U=C
Physical Unit for the temperature value
E=0.950
Emissivity setting
XG=1.000
Setting for transmission
For the calculation of the temperature value, it is possible to set an offset
(relative number to be added to the temperature value).
DO=-0.3
Offset adjustment -0.3 for the temperature signal
8.2.2
Post Processing
The following parameters can be set to determine the post-processing mode
P=5
maximum hold, hold time: 5 sec
F=12.5
minimum hold, hold time: 12.5 sec
G=10
averaging, average time (90%): 10 sec
8.3
Dynamic Data
To request the dynamic data, the following commands are available:
?T
Target temperature
?I
Detector ambient temperature
?XJ
Temperature of Thermocouple cold end (only valid for TC)
?Q
energy value of the target temperature
To check for resets (e.g. power shut down) use the command XI. Notice, after
a reset the unit is re- initialized.
?XI
asks for the reset status
!XI0
no reset occurred
!XI1
a reset occurred, new initialization of the unit
XI=0
sets the reset status back to 0