ASCII Programming
MI3
111
Description
Char
Format
P B S N Legal Values
Factory default Head /
Box
Restore Head Factory
defaults
HXF
nHXF
√
H
Head ambient
I
?nI
float
√
√
answer !nI=99.9
H
Switch panel lock
J
X
√
√
√
L = locked
U = unlocked
unlocked
B
Relay alarm output
control
depreciated for MI3 (use
the commands KB and
KH instead of)
K
K
integer
√
√
0 = off
1 = on
2 = Target norm. open
3 = Target norm. closed
4 = Intern norm. open
5 = Intern norm. closed
2
H
Relay alarm output
control
KB
KB
integer
√
√
0 = off
1 = on
2 = norm. open
3 = norm. closed
2
B
Relay alarm output
control
KH
nKH
integer
√
√
0 = off
1 = target temp.
2 = head ambient
1
H
Bottom temperature
value of output 1
L1O
(L)
float
√
√
√
L1O=
–40.0 or
L=
–40.0
0°C
B
Bottom temperature
value of output 2
L2O
float
√
√
√
L2O=
–40.0
0°C
B
Output 1 source
O1O
(O)
O1O=1T
float,
or nT,
or nI
√
√
v – float value
n – head number
if v = 60
– controlled by
head 1(O1O=1T)
O1O=0
B
Output 2 source
O2O
O2O=1I float,
or nT,
or nI
√
√
v – float value
n – head number
if v = 60
– controlled by
head 1(O2O=1I)
O2O=0
B
Peak hold time
P
nP
float
√
√
√
√
0.0 - 998.9s
(999 =
∞
)
0.0s
H
Power / AD value
Q
?nQ
H
Presel. Setpoint
SV
nSV
float
500°C
H
Target temperature
T
?nT
float
√
√
answer !nT=99.9
H
Get Analog Input 1
TV1I
?TV1I
0 – 5V
B
Get Analog Input 2
TV2I
?TV2I
0 – 5V
B
Temperature Unit
U
X
√
√
√
√
C / F
C
B/H
Poll / Burst mode
V
V=P
√
√
P=poll B=burst
poll mode
B
Burst string contents
X$
?X$
√
B
Multidrop Address
XA
0nn
√
√
√
000 – 032
(000 --> single unit mode)
000
B
1
setting average / peak / valley / advanced hold cancels all other hold modes
1.800.868.7495