Profinet
MI3
101
17.2.3
Input Data Structure
17.2.3.1
Station Input data
The input data length is 5 Byte.
Address without offset
Length
Format
Value
0
1 Byte
Bit0
Trigger state ( 0 – reset, 1 – set)
1
4 Byte
Real (Big Endian, Motorola)
Internal temperature of the station
17.2.3.2
Pyrometer Module Input Data
The input data length of pyrometer module is 8 Byte.
Address without offset
Length
Format
Value
0
4 Byte
Real (Big Endian, Motorola)
Object temperature
4
4 Byte
Real (Big Endian, Motorola)
Internal temperature
17.2.4
Output Data Structure
The pyrometer module provides output data. The output data length of pyrometer module is 3 Byte.
The output data may be used to change the initialization of the device (which was set once at start-up)
when the bus is in data exchange mode.
To do so the following structure is defined:
Address without offset
Length
Format
Value
0
1 Byte
Byte (Big Endian, Motorola)
Type of parameter
1, 2
2 Byte
Integer (Big Endian, Motorola)
Parameter
The <Type of parameter> gives the meaning of the following parameters (with the same format as
described in section 17.2.2.2
Number of type
Meaning
0
ignore output data
1
emissivity
2
laser
3
ambient temperature (for ambient temperature compensation)
4
averaging time
5
peak hold time
6
valley hold time
7
set point for the relay
If <Type of parameter> is set to 0 then the output data gets ignored. So it should be set to 0 as default.
17.2.5
Diagnostics
The diagnostics information of the fieldbus communicator can be read out acyclically using standard
diagnostics data sets defined in the Profinet IO specification.
Errors occurring when configuring and setting the parameters of the fieldbus communicator and the
connected pyrometer modules as well as external errors are reported by the communicator via
channel specific diagnostic.
In productive data exchange between the I/O controller and the fieldbus communicator, one byte IOPS
process data qualifiers are available for each module providing information of the validity of the
pyrometer module data (good/ bad). In the event of an error occurs during operation, the problem-