Operation
To operate the system, follow the steps below.
■
Turn on power supply.
■
Turn on the meter, chart recorder, or controller.
■
Read or monitor the temperature.
Warning
When using the air/water-cooled housing in a
heated environment run coolant continuously.
damage to the sensor and to the housing can
occur if coolant is turned off.
Maintenance
It is important to keep the sensor lens clean at all times to
ensure correct temperature measurement. If not using the air
purge accessory, blow off loose particles with clean
compressed air, then carefully wipe the surface with a
moist cotton swab (water or water-based glass cleaner).
dO NOT use solvents.
Troubleshooting
If your sensor is not performing as it should, use the table
below to match the symptom to the probable cause and solu-
tion.
Support
Our customer service representatives are always available to
provide application assistance, calibration, repair, and solutions
to specific questions or problems. Contact the Raytek Service
Department before returning any equipment. In many cases,
problems can be solved over the telephone.
For technical information refer to the CI Operator’s Manual.
© Raytek Corporation (3253679 Rev C1) 03/2011
Raytek and the Raytek logo are registered trademarks of Raytek Corporation.
Specifications subject to change without notice.
WARNING
INCORRECT WIRING WILL CAUSE SEVERE ANd
PERMANENT dAMAGE TO THE SENSOR.
Use the correct wiring chart to wire your sensor for
standard or high temperature cable wiring
CI sensor lens
www.raytek.com
The Worldwide Leader in Noncontact Temperature Measurement
Raytek Corporation
Worldwide Headquarters
Santa Cruz, CA USA
Tel:
1 800 227 8074 (USA and Canada, only)
1 831 458 3900
European Headquarters
Berlin, Germany
Tel:
49 30 4 78 00 80
China Headquarters
Beijing, China
Tel:
8610 6438 4691
To find a Raytek office near you, please visit www.raytek.com
Worldwide Service
Raytek offers services, including repair and calibration.
For more information, contact your local office or e-mail [email protected]
Symptom
Probable Cause
Solution
No Output
No power to sensor
Check power supply
Erroneous Temp Incorrect wire
Check wire color
connection
codes and reconnect
Erroneous Temp Faulty sensor cable
Verify cable continuity
Erroneous Temp Field of view
Remove obstruction
obstruction