Operation

To operate the system, follow the steps below.
 

 Turn on power supply.

 

 Turn on the meter, chart recorder, or controller.

 

 Read or monitor the temperature.

Warning

 When using the air/water-cooled housing in a  

heated environment run coolant continuously. 

damage to the sensor and to the housing can  

occur if coolant is turned off.

Maintenance

It is important to keep the sensor lens clean at all times to  
ensure correct temperature measurement.  If not using the air 
purge accessory, blow off loose particles with clean  
compressed air, then carefully wipe the surface with a  
moist cotton swab (water or water-based glass cleaner).  

dO NOT use solvents.

Troubleshooting

If your sensor is not performing as it should, use the table
below to match the symptom to the probable cause and solu-
tion.

Support

Our customer service representatives are always available to 
provide application assistance, calibration, repair, and solutions 
to specific questions or problems. Contact the Raytek Service 
Department before returning any equipment. In many cases, 
problems can be solved over the telephone. 
For technical information refer to the CI Operator’s Manual.

© Raytek Corporation (3253679 Rev C1) 03/2011 

Raytek and the Raytek logo are registered trademarks of Raytek Corporation. 

Specifications subject to change without notice.

WARNING

 

INCORRECT WIRING WILL CAUSE SEVERE ANd  

PERMANENT dAMAGE TO THE SENSOR.

Use the correct wiring chart to wire your sensor for  

standard or high temperature cable wiring

CI sensor lens

www.raytek.com

The Worldwide Leader in Noncontact Temperature Measurement

Raytek Corporation

Worldwide Headquarters

Santa Cruz, CA USA

Tel: 

1 800 227 8074 (USA and Canada, only)

 

1 831 458 3900

[email protected]

European Headquarters

Berlin, Germany 

Tel: 

49 30 4 78 00 80  

[email protected]

 

China Headquarters

Beijing, China

Tel: 

8610 6438 4691 

[email protected]

To find a Raytek office near you, please visit www.raytek.com

Worldwide Service

Raytek offers services, including repair and calibration.  
For more information, contact your local office or e-mail [email protected]

Symptom 

Probable Cause 

Solution

No Output 

No power to sensor 

Check power supply

Erroneous Temp  Incorrect wire 

Check wire color 

 

connection 

codes and reconnect

Erroneous Temp  Faulty sensor cable 

Verify cable continuity

Erroneous Temp  Field of view  

Remove obstruction 

 obstruction

Reviews: