5.3.1 Multiplexer Menu without Transformer Profiles
NOTE
⇒
Please see ‘90163-X.XX Instruction Manual WR-XX-Rack’ or ‘90164-
X.XX Instruction Manual WR-XX’ for instructions about switching the
use of transformer profiles on and off
When multiplexer menu (config) is opened as described in 5.3 Controlling a Multiplexer and
no profiles are used, the following screen appears:
When the button is touched as described above, a simple screen to control
Multiplexer with a dot matrix is opened. This is the way, to switch a winding resistance meter
to any test object. It’s recommended when the device under test is no transformer. The
following page shows how this screen looks, and how explains how to use it.
The radio buttons at the bottom of the screen must be set on WR to use the simple screen.
If a turns ratio meter should be connected to the test object, TR A:B or TR A:C must be
chosen. Please see chapter ‘5.5 Turns Ratio Measurement’ for details about turns ratio
measurement.
Channels and
allocated
Multiplexers are
listed here
Switch between
winding
resistance and
turns ratio mode
Touch to open
keyboard to
enter captions
Touch to open
Multiplexer
control screen