MAS: MEASUREMENT PARAMETER DEFINITIONS
mAs
is calculated from all recorded data.
mA
is calculated as an average of all samples above 50% of peak. Rush currents are automatically removed. Live
readings are based on samples since last reading. For long measurements, the final reading is recorded
approximately 1–2 s before end trig.
Time
starts the first time the current reaches 50% of the calculated mA and ends the last time it drops below 50%.
Live readings are time since start trig.
Pulses
are counted every time the signal reaches trig.
Pulse rate
and
mAs per pulse
are averages for the last 6 pulses.
PARAMETERS IN MA WAVEFORM
T
: Time
P1
,
P2
,
P3
: Pulses
mA
: mA
: Displayed mAs value.
NOTE!
If measurements are made simultaneously with a sensor and mAs cable, any shared
parameter (
time, pulses
or
pulse rate
) will be taken from the sensor.
RaySafe X2 User Manual
60