![RaySafe x2 User Manual Download Page 59](http://html2.mh-extra.com/html/raysafe/x2/x2_user-manual_3942726059.webp)
Switch on the X-ray generator.
Make an exposure and read the result.
NOTE!
Users of RaySafe X2 mAs must be aware of the potential damage to generators and of
electrical human hazards in case of improper connection or damaged equipment. RaySafe
X2 mAs is intended for use only by personnel authorized in performing calibration and
repair of X-ray equipment.
TIP!
You can measure with the R/F or MAM Sensor and the mAs cable both connected to get
radiation and generator current values at the same time.
TIP!
Tap a parameter for large digits, to get information about that specific parameter, and, if
applicable, see a waveform.
RaySafe X2 User Manual