RaySafe i3 Installation & service manual – Installation
3. Open the dosimeter options window by pressing
Dosimeter options
and then the
Settings
tab:
A
B
C
A. Verify that the computer time is correct and then synchronize the dosimeter time. Be sure to do this with all
dosimeters in a lab.
B. Set the
Real-time Display Mode
to
Show
, to make the dosimeter appear on a display.
C. Set
Power Mode
to
On
to start radio communication and dose logging.
NOTE:
Power Mode On
includes an automatic power saving function controlled by accelerometers. When
the dosimeter is stationary for 9 minutes, it switches to power save mode (no dose logging, no
radio communication). The dosimeter wakes up again when moved. If the dosimeter is stationary
positioned, not worn by a person, turn off the automatic power saving by selecting
Advanced:
Always on
.
4. Press
Information
:
A
B
B
C
A. Set
Full Name
and
Display Name
. The Display Name appears on the display. The Full Name is visible in the
data export files from the Real-time Hub and in Dose Viewer.
B. Change
Dosimeter Position
,
Clinical Role
and
Annual Dose Limit
if desired. These settings do not affect the
dose registration.