User’s guide Ytterbium Pulsed Fiber Laser RFL- P100M Ver.: 2.0
8
/ 12
Figure 6. Control sequence of laser.
T1
:
Represent the initialization time of the laser machine. The initialization time is less
than10s from power-on to perform initialization.
T2
:
Represent the pump recharge time, usually around 5ms.
T3
:
Represent the frequency switching time. Frequency needs to be adjusted before optical
laser output, and the switching time should not less than 5ms.
T4
:
Represent the pulse build-up time. Usually the build-up time is less than 200 us from the
pump current raising up to reach the 90% power of the optical pulse.
T5
:
Represent the pulse shut down time. Usually the shut down time is less than 100 us from
the pump current decliner to drop to the 10% power of the optical pulse.
6)
Definition of alarm signal
Table 5 Definition of alarm signal
PIN12
PIN11
PIN16
PIN21
Alarm content
×
Low
Low
Low
Laser Temperature
Alarm
1
×
Low
Low
High
Normal
×
High
Low
Low
System Error
2
×
Low
High
High
Power Error
3
PS 1
:
PIN 17 receives external DC voltage to provide power supply for alarm signal. Therefore,
please make sure PIN17 is connected to the +5V
power supply. The temperature warning will
remain if the power or DB25 port is not well connected.