ELECTRIC MOTOR

 

The motor installed on Foster model trimmers is equipped with permanently 
lubricated, sealed ball bearings. No lubrication or maintenance is required. 

 

DOOR SECURE 

When closing and securing the door, do NOT tighten the door secure knob. It is 
only necessary for the door to "touch" the base casting for a leak free fit.  Over 
tightening may cause leakage. 

 

USE PLENTY OF WATER 

The use of an adequate amount of water while trimming will keep the wheel 
clean and cutting. An inadequate volume of water will result in a clogged 
unusable wheel. 

 

CHECK THE WATER SPRAY TUBE 

It's a good idea to occasionally check the water spray tube to be sure all holes 
are open. A straighten paper clip is useful in opening plugged holes. 

 

WATER VALVE

 

The brass water valve located on the door of the model trimmer is intended to be 
used as a flow control only, and not intended to be used as a water shut-off 
valve. 

 

MODEL AND SERIAL NUMBER 

When ordering parts or requesting service always give the machine model 
number and serial number which are inscribed on the identification plate located 
on the front of the Lab Master Model Trimmer. 
 
 

EXCESSIVE VIBRATION 

 

Excess vibration from your model trimmer is a result of one or both of the 
following: 

1.  An imbalance in the aluminum back plate as a result of accumulation of 

plaster and stone on the back.  

2.  An out-of-balance abrasive wheel due to stone build-up and/or wear, or an 

improperly balanced or non-balanced wheel.  NOTE:  Abrasive wheels 
purchase from us are precision balanced for smooth running. This is done by 
removing a portion of the heavy edge. As a result of the balancing process 
the wheel may be out-of-round, but is in balance. 

 

To determine and correct an imbalance follow this procedure: 

1.  Remove the abrasive wheel from the model trimmer, close and secure the 

door, and turn the model trimmer on with the wheel removed. If the vibration 
has ceased, the problem lies with the grinding wheel, and a replacement is 
needed. NOTE:  Sometimes the center mounting devise of the wheel (P/N 
M095 - Lock Button) is worn and does not center the wheel properly. A wheel 
that is mounted off center will also cause vibration. To check the Lock Button, 
remove it together with the wheel, and check it's fit in the center hole of the 

Summary of Contents for MT10

Page 1: ... MT12 and MT15 Operating and Maintenance Instructions Sales Service and Technical Assistance Call Toll Free 800 654 4519 Ray Foster Dental Equipment 5421 Commercial Drive Huntington Beach CA 92649 714 897 7795 Fax 714 898 6013 ...

Page 2: ...the country of destination Use only this power cord or one that is approved for use in your country Plug into a Branch Circuit with 16A Circuit Breaker To reduce the risk of fire and to protect the units electrical system plug into a branch circuit with a 16 amp circuit breaker Disconnect the Power Cord Before performing any maintenance or repairs turn the unit off and unplug the power cord to pre...

Page 3: ... complete the installation A local plumbing supply will usually have the required adapter NOTE It is recommended that you do NOT use copper tubing to connect the model trimmer to the water supply The problem with copper tubing is that it is not flexible This nullifies the feature of the hinged door because the hose must flex when opening the door Water is discharge from the model trimmer through a...

Page 4: ...hich are inscribed on the identification plate located on the front of the Lab Master Model Trimmer EXCESSIVE VIBRATION Excess vibration from your model trimmer is a result of one or both of the following 1 An imbalance in the aluminum back plate as a result of accumulation of plaster and stone on the back 2 An out of balance abrasive wheel due to stone build up and or wear or an improperly balanc...

Page 5: ...N M096 Cork Gasket C from the backside of the attach button and replace it with the new gasket included with all Foster replacement wheels If only reversing the existing wheel or using a wheel from a source other that Ray Foster make sure that the existing cork gasket is in good condition if not a new one is needed This is essential in preventing slippage of the abrasive wheel 3 Install the new re...

Page 6: ... your 12 Foster Model Trimmer to an Orthodontic Model Trimmer capable of trimming accurate and attractive study models Back Plate Removal Tool Cat No M816 If the Aluminum Back Plate P N M221 M222 must be removed for imbalance corrections repairs etc a Back Plate Removal Tool is essential Simple illustrated instructions included Splash Shield Cat No M005 Protect your face and clothing from splatter...

Page 7: ...PARTS DIAGRAM ...

Page 8: ...M T sold as set only with Base Casting M241 Door for 10 M T sold as set only with Base Casting 18 M018 Water Spray Tube Valve Assembly M019 Water Spray Tube Valve Assembly 19 M244 Plastic Knob Tray Secure 20 M245 Washer Nylon 21 M117 Water Supply Hose 3 Feet 22 M108 Door Lock Bolt 23 M095 Lock Button with Screws Gasket 24 M096 Lock Button Gasket 25 M247 Standard Work Tray for 10 M T M248 Standard ...

Page 9: ...rt together with proof of date of purchase must be delivered freight prepaid to the factory in Huntington Beach California At its option Ray Foster Dental Equipment Company will replace or repair the defect free of charge providing the Model Trimmer has not been subjected to abuse neglect damage or used in a capacity other than which was intended Ray Foster Dental Equipment 5421 Commercial Dr Hunt...

Reviews: