Operating the Rotary Suspension Mixer

 

 

 

 

CAUTION 

When working with glass tubes, flasks, bottles, breakable or other brittle or fragile articles it is important to wear appropriate 
safety eyewear and protective gloves to ensure operator safety. In addition, the Rotary Suspension Mixer should be used in 
accordance with your operational health and safety policies. 

 

 

 

-

 

Load your tubes or bottles evenly around the disc to present a balanced load to the mixer 

-

 

Ensure that there are no obstructions which would prevent proper rotation of the disc. 

-

 

Adjust the tilt to the desired angle. 

-

 

Switch on the power. 

-

 

Select the desired agitation speed. Refer to the section in this User Manual titled “Adjusting The Mixing Control” for 
instructions. 

-

 

The speed may be adjusted as desired. 

-

 

Once desired mixing has been achieved turn the speed dial anti-clockwise to slow the disc. 

-

 

Remove the tubes or bottles once mixing is complete. 

-

 

If the mixer is no longer being used, switch the power to Off (element 1 in Figure 1) 

 

 

 

Storing & Relocating The Rotary Suspension Mixer

 

 

 

 

The Rotary Suspension Mixer should be stored out of direct sunlight at an ambient temperature below 30° Celsius in a clean 
and dry location which meets the environmental conditions required as detailed in the technical specifications of this User 
Manual. 
 

-

 

Switch the power to “Off” (element 1 in Figure 1) 

 

-

 

Unplug the equipment from the mains power supply outlet. 

 

-

 

Ensure the Rotary Suspension Mixer is clean and dry. If required, clean it according to the section of this User Manual titled 
“Routine Cleaning & Maintenance”. 

 

-

 

Relocate it to an appropriate storage location. Ensure the Rotary Suspension Mixer is stored in a clean and dry location 
away from potential damage by accidental knocks and bumps. 

 
 

Summary of Contents for RSM7DC

Page 1: ...5 06 2015 English www ratek com au 60 Wadhurst Dve Boronia Victoria 3155 Australia P 613 9887 2161 F 613 9887 2163 E sales ratek com au All Rights Reserved Made in Australia User Manual printed in Aus...

Page 2: ...roduct to ensure it has not been damaged in transit Any damage should be reported immediately to the responsible carrier If the product is damaged in any way re pack the product into the supplied pack...

Page 3: ...PERATOR RESPONSIBILITY SAFETY CONSIDERATIONS 4 4 4 4 Conditions of Operation 4 Safety Labels And Markings 5 OPERATING PROCEDURES 6 6 6 6 Identification of Controls Functions 6 Safety Warnings 7 Safety...

Page 4: ...e consisting of a power switch and a mixing speed control knob These controls allow the operator to set a required agitation speed The Rotary Suspension Mixer should be operated strictly in accordance...

Page 5: ...ventilated areas The unit is not intended for use in explosive atmospheres or in confined spaces The allowed operating environment must be between 5 Celsius and 40 Celsius ambient air temperature The...

Page 6: ...ainable user safety Identification of Controls Functions Throughout this User Manual the numerical element representing a control or component of the Rotary Suspension Mixer will be used to identify i...

Page 7: ...lity as well as providing some protection against lightning strikes The equipment should be operated on a good reliable supply of power at all times Note The Rotary Suspension Mixer should not be oper...

Page 8: ...ct or wall Ensure that there is a suitable mains power supply outlet within reach of the supplied mains power supply without placing any strain whatsoever on the lead or plug The Rotary Suspension Mix...

Page 9: ...e If you are unsure about the rating of your power supply consult your workplace administrator or electrical maintenance staff to determine if your power supply is suitable for use with this product b...

Page 10: ...d may be adjusted as desired Once desired mixing has been achieved turn the speed dial anti clockwise to slow the disc Remove the tubes or bottles once mixing is complete If the mixer is no longer bei...

Page 11: ...e substances that may compromise the integrity or function of electrical insulation electrical components or other components Ratek are consulted prior to decontamination or cleaning being undertaken...

Page 12: ...r a table of factory fitted warning labels If any labels are missing illegible or otherwise not functional contact your supplier to obtain new replacement labels before operating the equipment Ensure...

Page 13: ...o 50 relative humidity at 40 Celsius MAINS supply voltage fluctuations up to 10 of the nominal voltage Over voltage Category II Pollution Degree 2 Disposal At end of life this equipment should be disp...

Page 14: ...oting The Rotary Suspension Mixer provides a simple to operate user interface when used in conjunction with this User Manual If at any stage you experience abnormal operation anything other than that...

Page 15: ...ent from Ratek Instruments For a period of 12 months from date of purchase service labour and repairs will be carried out at no charge by an approved repairer or Ratek Instruments at the discretion of...

Page 16: ...ferred to a local repair agent for service Clean the unit thoroughly in accordance with this Operating Manual If necessary decontaminate the unit to ensure safety for the service technicians Pack the...