21 

Kestrel/USER-MANUAL/REV1.1/08-20 

10.2 Contrast Modification (XCAP Std. Only) 

The image contrast can be modified from the 

“Contrast Modification”

 section under the 

“Modify”

 tab in the XCAP imaging window. Where to find this control feature is shown in 

Figure 13. If wanting to view a useable image without any EM gain applied, contrast 
modification will have to be enabled. Raptor recommends having this enabled at all times 
when imaging the camera. If applying ROI or binning controls, the contrast modification will 
have to be re-enabled each time.

 

 

 

 

 

 

 

 

 

 

 

 

 

In the contrast modification box, that can be seen from Figure 14, select 

“Stretch Contrast, 

Histogram Percentile Endpoints” 

and click 

“preview”.

 The contrast modification will now be 

applied over the live image feed. The contrast can be adjusted using the low and high end 
percentile point controls. The default settings are usually adequate for most applications. 

NOTE: This control feature is only applicable with XCAP Std. 

 

 

 

 

 

 

 

 

 

 

Figure 14: Contrast Modification. 

Figure 13: Contrast Modification Location on Toolbar. 

Summary of Contents for KE60V-BV-CL

Page 1: ...1 Kestrel USER MANUAL REV1 1 08 20 Kestrel Model KE60V BV CL USER MANUAL...

Page 2: ...CHILLER SETUP 9 6 1 Connecting Camera to Frame Grabber 9 6 2 Connecting Camera to Chiller 9 6 3 Recommended Coolants for Chiller 9 6 4 Setting the Coolant Temperature for Re circulation 9 6 5 Drainin...

Page 3: ...USER MANUAL REV1 1 08 20 9 5 Maximum Frame Rates 18 9 6 Thermoelectric Cooler TEC 18 9 7 Information Tab 19 10 XCAP CONTROL FEATURES 20 10 1 Saving Preset Settings 20 10 2 Contrast Modification XCAP S...

Page 4: ...and explained with the use of XCAP Imaging software which is the core plug and play software package that is offered with Raptor cameras An image of the camera is shown in Figure 1 2 CAMERA CARE 2 1...

Page 5: ...eed low noise electronics provide linear response and sensitivity for rapid image capture The Camera Link digital interface provides the most stable platform for data transfer and the camera will work...

Page 6: ...estrel USER MANUAL REV1 1 08 20 4 DESIGN OVERVIEW 4 1 Mechanical Model PDF of mechanical model available from our website https www raptorphotonics com products kestrel emccd Units shown in mm and inc...

Page 7: ...or Optical Table The camera has a 20 BSW Whitworth threaded hole to mount to a tripod or an optical table 2 3 4 1 6 1 Label Recess 2 3M Camera Link connector Part 12226 1150 00FR 3 SMA connector Trigg...

Page 8: ...n request Raptor can create a LabView icd file which can be used to control the camera on National Instruments imaging tool NI MAX The file may also be useful if attempting to create your own LabView...

Page 9: ...free path for flow of air entry and exit at the left side and right side of the T257P chiller prior to operation 4 Remove the reservoir cap and add coolant to the reservoir until the fluid reaches th...

Page 10: ...sed on the amount of use you may need to perform this more frequently For more detailed information please refer to the Thermotek user manual If using another model of chiller please refer to its user...

Page 11: ...CI EB1 providing roughly 200MB sec maximum bandwidth Contact EPIX Inc for further information regarding minimum computer laptop specification requirements to run the XCAP Imaging Software http www epi...

Page 12: ...Pre release version with support for the latest cameras and latest PIXCI imaging boards Open the downloaded file when complete and follow the onscreen instructions in the installation wizard If a pop...

Page 13: ...all control parameters set to the default states Open w Last used Video Setup will open the control panel with all parameters set at the last known state Once this option between the two has been sele...

Page 14: ...ou are obtaining video data from the camera The imaging statistics displayed directly underneath the imaging window will also inform you if you are receiving a video feed from the camera Once you have...

Page 15: ...ling the amount of voltage that the EM gain register in the camera is seeing The relationship of gain to the DAC value is exponential so up until around 3000 on the slider there is a minimal gain bein...

Page 16: ...camera into external trigger mode Enabling the external trigger will open the trigger polarity options rising or falling edge There is a slight drop off in the maximum frame rate in external trigger m...

Page 17: ...x2 and 4x4 9 4 Exposure Time and Frame Period The exposure time and frame period controls are active in the control panel under all tabs except the port and info tabs Unless Live FFR mode discussed in...

Page 18: ...ling is 20 C If only using the fan the minimum cooling temperature is 20 C For important information on using the liquid cooling with the chiller please refer to sections 6 2 6 5 The actual TEC set po...

Page 19: ...Information Tab The Info tab displays the manufacturers data of the camera such as the firmware version serial number etc The sensor and PCB temperature can also be read from this tab The Info tab is...

Page 20: ...frame grabber settings can be saved in the Preset tab under the PIXCI relevant frame grabber model number section of the GUI as shown in Figure 12 Up to three different presets can be saved per settin...

Page 21: ...en imaging the camera If applying ROI or binning controls the contrast modification will have to be re enabled each time In the contrast modification box that can be seen from Figure 14 select Stretch...

Page 22: ...22 Kestrel USER MANUAL REV1 1 08 20 CORPORATE HEADQUARTERS Raptor Photonics LTD Willowbank Business Park Larne Co Antrim BT40 2SF Northern Ireland PH 44 2828 270141 www raptorphotonics com...

Reviews: