5.2.2 Connecting the Camera and Chiller 

Raptor uses a Chiller from Thermotek, the T257P Precision Chiller. For the datasheet and 
full user manual from Thermotek, please contact Raptor and we can provide this. The 
manual should be included with the chiller, however. The instructions to set up the chiller are 
as follows: 

1. Connect the T257P chiller to the Eagle XO using the tubing provided. Connection to the 
coolant channel on the camera is made via two quick release valve coupling bodies. The 
polarity of the tubing connections does not matter. 

2. Keep the chiller horizontal and on a level surface.  

3. 

Make sure there is a minimum 12” clearance and free path for flow of air entry and exit at 

the left side and right side of the T257P chiller prior to operation.  

4. Remove the reservoir cap and add coolant to the reservoir until the fluid reaches the 
bottom of the neck. Please refer to section 6.3 for recommended coolants.  

5. Close the cap securely to the reservoir. Make sure not to overfill the reservoir.  

6. Install the appropriate end of the power cord into the unit and connect to the mains 
electricity. 

 

If using another chiller model, please refer to its manual for start-up instructions. 

 

5.1.3 Recommended Coolants for the Chiller 

The recommended coolants are: 
  

Option 1:

 Distilled Water  

Option 2:

 95% distilled water and 5% isopropyl alcohol mixture prevents bio growth.  

Option 3:

 80% distilled water and 20% inhibited Glycol mixture for set temperatures below   

5

 

 ͦ

C. 

 

 

Raptor recommends using option 3. If purchasing the Thermotek model from Raptor or using 
a demo kit provided by Raptor, there will be inhibited Glycol mixture provided to mix with 
water. 

 

5.1.4 Setting the Coolant Temperature for Re-circulation 

When powering up the chiller, the screen will immediately give the option to set the 
temperature of the coolant. It is recommended to set a 20

 

 ͦ

C set point. After the set point has 

been configured, selecting start will initiate the coolant re-circulation. For more detailed 
information on all the operating modes of the T257P Precision Chiller, please refer to the 
Thermotek user manual.

 

Never cool the sensor below the dew point of the environment that it is in, otherwise 
permanent damage due to condensation may occur. Switch the cooling off 
immediately if you suspect condensation is forming on the sensor. 

 

5.2 Connecting the Camera and Frame Grabber 

The camera has the normal MDR port on the interface. The main frame grabber that Raptor 
offer with this camera has the normal MDR port (EPIX EB1). Therefore, one MDR 

– MDR 

Camera Link cable is required to connect the camera to this frame grabber.

 

Summary of Contents for Eagle EA4240XV-BN-CL

Page 1: ...1 Eagle XV Models EA4240XV BN CL EA4240XV BNDD CL EA4710XV BN CL EA4710XV BNDD CL USER MANUAL...

Page 2: ...um Pressure 11 5 CAMERA CHILLER SETUP 12 5 1 Liquid Cooling 12 5 1 1 Liquid Cooling Connection to the Camera Head 12 5 2 2 Connecting the Camera and Chiller 13 5 1 3 Recommended Coolants for the Chill...

Page 3: ...3 Trigger Control 23 9 3 1 Live ITR Integrate then Read 23 9 3 2 Live FFR Fixed Frame Rate 23 9 3 3 Ext Triggered External Trigger 23 9 3 4 Btn Triggered Snapshot Trigger 23 9 3 5 Pixel Readout Clock...

Page 4: ...pe Detailed information is provided on each of the cameras control parameters as well as stating important precautions to take when operating the camera Each camera control is discussed and explained...

Page 5: ...and liquid cooling in a vacuum environment This level of cooling gives an ultra low dark current reading of 0 0005e p s in the standard sensor format allowing for longer integration times The Camera L...

Page 6: ...Mechanical Model Power supply weight 1 5kg Units shown in inches mm Figure 2 Solid Works Model Basic 2D mechanical dimensions of the camera PDF of the 2D mechanical model available from our website w...

Page 7: ...7 Figure 3 Solid Works Model basic dimensions of the power supply module Power supply weight 0 5kg Units shown in inches mm...

Page 8: ...ng for O D pipe 3 TTL Trigger output 50ohm SMA Exposure 4 TTL Trigger output 50ohm SMA Readout 5 Camera link connector Base MDR 3M P N 10226 6212PC 6 Label recess Model Number Serial Number 7 TTL Trig...

Page 9: ...g by screwing the module to a metal surface 3 5 Trigger IN External synchronisation with the start of integration signal may be achieved using the Trigger IN connector Input impedance 510 200pF input...

Page 10: ...tamination from the local environment see Figure 4 It is recommended to keep the temporary plate and O ring for storage of the camera when not in use or for transportation Regarding storage temperatur...

Page 11: ...ensor if this background signal is affecting your data acquisition it may be necessary to switch off the pressure gauge once you are satisfied that steady state conditions have been achieved Never coo...

Page 12: ...rwise damage to the camera will occur 5 1 1 Liquid Cooling Connection to the Camera Head The camera electronics drive a ThermoElectric Cooler TEC to cool the CCD The hotside of the TEC MUST be cooled...

Page 13: ...Recommended Coolants for the Chiller The recommended coolants are Option 1 Distilled Water Option 2 95 distilled water and 5 isopropyl alcohol mixture prevents bio growth Option 3 80 distilled water...

Page 14: ...e of the feedthrough must protrude into the vacuum chamber The power cable with grey insulation is intended for in vacuum connection Always ensure the cable polarity is correct before attempting to pl...

Page 15: ...ended cut out is 65mm x 26mm as shown in Figure 8 The bulkhead flange must be secured using 4 off screws at the corners also indicated in Figure 8 Figure 8 Dimensions and cut out for Camera Link Feedt...

Page 16: ...e and format options 6 4 Liquid Feedthrough Optional Item Coolant connections to the Swagelok fittings on the camera head can be made using OD pipe and a feedthrough flange such as that shown in Figur...

Page 17: ...ile which can be used to control the camera on National Instruments imaging tools such as NI MAX The file may also be useful if attempting to create your own LabView VI 7 4 Custom Software Interfacing...

Page 18: ...rom another company the specification requirements of this frame grabber must meet those supplied by the PIXCI EB1 model If using a laptop EPIX offers base Camera Link frame grabbers for a laptop syst...

Page 19: ...ing Open w Default Video Setup will open the control panel with all control parameters set to the default states Open w Last used Video Setup will open the control panel with all parameters set at the...

Page 20: ...l have three moving dots that flash This indicates that you are obtaining video data from the camera The imaging statistics displayed directly underneath the imaging window will also inform you if you...

Page 21: ...p gain will be enabled offering the lowest noise If you are imaging with a higher intensity signal high preamp gain can be disabled The exposure time and frame rate controls are shown under every tab...

Page 22: ...closed Permanently open Open for the duration of a valid exposure time Additional delay registers may also be programmed to allow a shutter to be fully open before and fully closed after the exposure...

Page 23: ...osure Rate becomes active when this mode is selected giving the user control to program the frame rate 9 3 3 Ext Triggered External Trigger If this mode is enabled then the camera will use an external...

Page 24: ...mage may be read from the camera instead of CCD data The test image will consist of a fixed ramp pattern that will start with a value of 0 on the first pixel read from the camera and increment by one...

Page 25: ...gle XV 47 10 model Also ROI width and ROI height must be 0 Binning In addition to standard 1 1 output various levels of pixel binning may be programmed up to 64 64 Pixel binning is performed on the CC...

Page 26: ...ing must be used when using the TEC This is needed to dissipate heat away from the hot side of the TEC Do not enable the TEC if liquid cooling is not being used otherwise damage to the camera will occ...

Page 27: ...the camera was built ADC DAC Calibration Values These are calibration values that are used to set the TEC set point and read back the sensor and PCB temperatures They are not needed for the user as t...

Page 28: ...ixinc com manuals pixci_eb1 index htm This section will discuss in detail a few features on XCAP that Raptor thinks would be of immediate use when using the camera 10 1 Recording Images on XCAP Captur...

Page 29: ...settings file If the camera is set to a desired state outside of the default state clicking Save 1 will save all the current parameter states that have been set This can be done a further two times Th...

Page 30: ...ication will have to be re enabled each time In the contrast modification box that can be seen from Figure 24 select Stretch Contrast Histogram Percentile Endpoints and click preview The contrast modi...

Page 31: ...31 CORPORATE HEADQUARTERS Raptor Photonics LTD Willowbank Business Park Larne Co Antrim BT40 2SF Northern Ireland PH 44 2828 270141 www raptorphotonics com...

Reviews: