5. Insert a new sample filter.
6. Rotate the filter housing clockwise.
7. Replace the filter cover and four screws.
Power Up the Device
Press the power button so the device is
ON, and allow it to warm up.
Burn in the Membrane
1. Log onto the ITMS™ software.
2. Burn in the membrane.
see "Burn In Membrane" on page 79)
Perform a Calibration and Verification
1. Perform a calibration and verification.
see "Calibration and Verification" on page 23)
2. Ensure there are no alarms or warnings.
Return to Operation
Make all entries in your device’s maintenance log as directed by your administrator.
As-Needed Maintenance
To keep the device operating at peak performance, the procedures listed in this section should be performed as-
needed.
CAUTION:
When maintaining the device, you must wear clean cotton gloves and use only Rapiscan
Systems-supplied consumables, parts, and supplies.
NOTE:
The ability to perform some maintenance procedures may depend upon the privilege levels
assigned.
CHAPTER 7 │User's Guide
127
Rapiscan Systems Proprietary and Confidential
Summary of Contents for Mobile Trace
Page 1: ...MobileTrace User s Guide For P0007027 MA001133 Revision ...
Page 2: ......
Page 31: ...This page intentionally left blank to ensure new chapters start on right odd number pages ...
Page 60: ...CHAPTER 4 Menu Functions Main Screen 44 Menu Functions 45 ...
Page 164: ...CHAPTER 9 Parts and Accessories Device and Replacement Part Numbers 148 ...
Page 167: ...This page intentionally left blank to ensure new chapters start on right odd number pages ...
Page 178: ...For P0007027 MA001133 Revision C ...