Vulcan® three

-

stage burnout furnace

 

Select one of the programs by pressing the knob. The arrow will stop blinking to indicate that the program has been 
selected. 
 
Each program has three stages that allow the following parameters to be set: 

• 

RMP: ramp temperature rise per minute 

• 

TMP: desired temperature to be reached in the stage 

• 

HLD: the length of time you want the set temperature to be held 

 
Once the parameters in each stage are reached, the unit will move to the next stage. 
 
To change parameters, rotate the knob to move the arrow to the desired parameter. Press the knob to select the 
parameter. Turn the knob to change the value. Press the knob to save the value. Continue this process until all of 
the values have been set. Rotate the knob to move the arrow to SAVE at the bottom of the screen and press the knob 
to save the program. Rotate the knob to move the arrow to EXIT and press knob. You will be returned to the START/
STOP SYSTEM screen. 
 
To create additional programs, press the knob to return to the RAMP/SOAK PROGRAMS and follow previous 
procedure. 
 

4.5 

Starting the Oven 

If you are not on the START/STOP SYSTEM screen, turn the knob to move the arrow to EXIT and press the knob. When 
the START/STOP SYSTEM screen is displayed, the arrow should be aligned with START/STOP SYSTEM. 

 

 
Press the knob to move to the STATUS: HEATING SCREEN. 

 

 
There are four parameters on this screen and two start options: 

• 

MODE: choice of nine three-stage programs and one fixed temperature 

• 

LINK: allows you to link two three-stage programs, creating six stages 

• 

DELAY: permits setting a delayed start time 

• 

END: can set to turn off or hold temperature, with or without audible 

 

Summary of Contents for Vulcan

Page 1: ...ulevard PO Box 1570 Maumee OH 43537 USA 800 253 4502 419 865 9497 419 865 9997 FAX www ransom randolph com Investing with Innovation Made in the USA Ransom Randolph Instruction Manual Vulcan three sta...

Page 2: ...ntenance Error Messages 14 15 6 Accessories and Replacement Parts 15 7 Disposal 16 8 Product Service 16 17 9 Warranty 17 Copyright This instruction manual is protected by copyright and is intended exc...

Page 3: ...ill lead to expire the warranty granted by Ransom Randolph R R Examples of conceivable improper uses Firing of unapproved materials Unsupervised operation Operation by non qualified personnel 1 2 Oper...

Page 4: ...upervision of qualified personnel Definition of the responsibilities of staff members for commissioning operation cleaning troubleshooting and maintenance Verified availability legibility and complete...

Page 5: ...state and federal regulations This product and many similar products on the market today contain crystalline silica and ceramic fibers In addition there are thread lockers that contain a small amount...

Page 6: ...the unit to a protected main outlet according to the technical data on page 8 The electrical service provided must be a dedicated line of the proper size according to local electrical codes Do not use...

Page 7: ...ation 7 Made in the USA Vulcan three stage burnout furnace 1 9 Description of Unit 1 Oven Door 2 Display 3 Control Knob 4 ON OFF Switch 5 Circuit Breaker 6 IEC Socket Main Power 7 Program Parameter LE...

Page 8: ...0003 5355740004 Volts 200 240 V 100 120 V 200 240 V 100 120 V Amps 4 4 A 12 A 10 A 14 7 16 0 A Watts 1060 W 1440 W 2400 W 1470 1920 W Watts to Maintain 1832 F 1000 C Air Exchange setting at 0 525 525...

Page 9: ...elves and other airflow restrictions Connect the line cord packaged in the rear of the furnace to the socket in rear of furnace Plug the other end into a properly grounded outlet Combustion Burnout or...

Page 10: ...iagnostic test The display shows the Remove Packing reminder and alternately the Click To Continue command Remove the packaging material if not done so already and push the adjustment wheel The unit i...

Page 11: ...set the parameters rotate the knob to move the arrow to the desired parameter Press the knob to select the parameter Rotate the knob to change parameter Press the knob to save parameter Continue this...

Page 12: ...ress the knob to save the value Continue this process until all of the values have been set Rotate the knob to move the arrow to SAVE at the bottom of the screen and press the knob to save the program...

Page 13: ...combine programs or have FIXED TEMP selected for MODE DELAY this parameter is used to set delayed start Rotate knob to move arrow to DELAY Press knob to select Rotate knob to change start hour press k...

Page 14: ...er normal use the furnace should not require calibration The electronics used in the Vulcan furnaces are very stable and will have minimal drifts over the life of the furnace Thermocouple replacement...

Page 15: ...rror Code 4 Backwards thermocouple Call R R Service Department Displays Error Code 8 Circuit board over temperature Call R R Service Department Displays Error Code 16 The muffle temperature exceeded 1...

Page 16: ...e packaging materials in an environmentally responsible manner separated by materials 8 Product Service Three methods of product service are available Telephone assistance available at the number list...