Calibration
The 76634-00 Multi-Function Meter is not field
repairable. The meter should be returned to the
factory for repair or recalibration.
Ransburg recommends that the Multi-Function
Meter and its associated equipment (Paint probe,
kV Probe, and/or Test Leads) be sent in together
for recalibration at least once per year.
The best recalibration results the
Multi-Function Meter and accesories
(Paint probe, kV Probe, and/or Test
Leads) should be sent in together for
recalibration.
When shipped from the factory, and
after each calibration, the Multi-Function
Meter will be sealed with a calibration
sticker noting the next date when
calibration expires. Removal of this
seal will void calibration of the unit.
A certificate of calibration will also be
shipped with each calibrated unit.
NOTE
GENERAL MAINTENANCE
Maintenance - Battery Replacement
When required, the battery will need to be replaced.
It may be purchased locally and is a standard
alkaline type 9V battery. To replace the battery,
perform the following steps (refer to Figure 11).
1. Remove the sliding back cover of the meter.
2. Remove the existing battery from the two (2)
prong 9V connector.
3. Replace with new alkaline 9V battery.
4. Replace back cover.
5. Appropriately discard used battery.
The screen will go blank when the
battery needs to be replaced.
NOTE
Figure 11: Battery Replacement
Test Equipment - Introduction, Operation & Maintenance
TE-98-01.7