Basic Operation
R&S
®
ZVA-Z75, -Z90, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500
19
Quick Start Guide 1307.7039.62 ─ 10
2. Connection of the frequency converters
3. Calibration using a suitable waveguide calibration kit
4. Connection of the DUT and measurement
3.3 Activating the Frequency Converter Mode
To activate the converter mode for a setup without external generator,
1. Click "System > System Config ..." and open the "Frequency Converter" tab of the
"System Configuration" dialog.
2. Select your converter "Type": R&S
ZVA-Z75, R&S
ZVA-Z90, R&S ZVA-Z110,
R&S
ZVA-Z140, R&S
ZVA-Z170, R&S
ZVA-Z220, R&S
ZVA-Z325, or R&S
ZVA-
Z500.
3. Select a test setup with an analyzer port as external source, click "Apply" to acti-
vate the frequency converter mode and "Close".
Analyzer settings with active frequency converter
In frequency converter mode, the frequency and level settings of the network analyzer
are automatically set to be compatible with the selected frequency converters. "Low
Phase Noise" is enabled, Automatic Level Control (ALC) is disabled. The frequency
and levels of all ports are displayed in the "Port Configuration" dialog ("Channel >
Mode > Port Config …").
Activating the Frequency Converter Mode