User Manual 1178.5966.02 ─ 19
[SENSe:]POWer...
The
[SENSe:]POWer...
commands configure the optional receiver step attenuators
Chapter 9.7.6, "Receiver step attenuators"
..................................................................................949
[SENSe<Ch>:]POWer:ATTenuation
<arg0>[, <arg1>]
Sets an attenuation factor for the received waves. This command is available if at least
one of the
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0>
Physical port number
Instead of port numbers 1 and 2 you can also use the constants
ARECeiver and BRECeiver, respectively.
Default unit: n/a
<arg1>
Attenuation factor for the received wave.
Range:
0 dB, 10 dB, 20 dB, 30 dB. The analyzer rounds
any entered value below the maximum attenuation
to the closest step.
*RST:
0 dB
Default unit: dB
Example:
POW:ATT AREC, 10
Set an attenuation factor of 10 dB for the waves received at test
port 1 and channel no. 1. The other test ports and channels are
not affected.
Manual operation:
See
[SENSe:]SEGMent...
The
[SENSe:]SEGMent<Seg>...
commands define all channel settings for a seg-
mented frequency sweep. A segmented sweep is activated via
SEGMent
.
The commands in this subsystem do not accept the step parameters UP and DOWN.
Numeric values can be entered directly or using the DEFault, MINimum, MAXimum
parameters.
................................................................................950
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
........................................................ 950
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
...........................................951
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect
............................................. 952
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol
................................952
....................................................................................953
[SENSe<Ch>:]SEGMent<Seg>:DEFine
............................................................................954
VNA command reference