DISPlay<1...2>[:WINDow]:TRACe<1...2>:MEMory<1...3>[:STATe]
●
DISPlay<1...2>[:WINDow]:TRACe<1...2>:MODE
●
[SENSe:]DETector<1...2>[:FUNCtion]
●
[SENSe:]DETector<1...2>[:FUNCtion]:AUTO
●
●
●
●
●
MEASurement<1...2>:ISUP:APER:AUTO
●
For a detailed description of the commands not described below, see
on page 248 in the CAT mode.
DISPlay:LAYout
<LayoutMode>
This command arranges the diagrams in the screen, leaving the diagram contents
unchanged.
Parameters:
<LayoutMode>
SINGle | DHORizontal | DVERtical | TRIPle | QUADruple
SINGle
One measurement trace window is displayed.
DHORizontal
Two measurement trace windows are arranged in top and bot-
tom rows.
DVERtical
Two measurement trace windows are arranged side by side.
Triple
Three measurement trace windows are displayed with one at the
top and two at the bottom arranged side by side.
QUADruple
Four measurement trace windows are displayed with two at the
top and two at the bottom rows.
Example:
DISP:LAY SING
Single measurement trace window is displayed.
[SENSe:]DETector<1...2>[:FUNCtion]
<Detector>
This command selects a detector function.
The numeric suffix at DETector specifies the primary or secondary detector.
Suffix:
<1...2>
.
1...2