Putting the Analyzer into Operation
R&S
®
ZNA
12
Getting Started 1178.6456.02 ─ 05
Packing material
Retain the original packing material. If the instrument needs to be transported or ship-
ped later, you can use the material to protect the control elements and connectors.
Risk of injury during transportation
The carrying handles at the front and side of the casing are designed to lift or carry the
instrument. Do not apply excessive force to the handles. If a handle is ripped off, the
falling instrument can cause injury.
As the R&S
ZNA
is very heavy
(over 30
kg fully equipped), it must always be carried
by two people using both carrying handles to avoid personal injury or damage to the
instrument.
3.2
Positioning the Instrument
The network analyzer is designed for use under laboratory conditions, either on a
bench top or in a rack. Notice the general ambient conditions at the operating site
described under
"Risk of instrument damage due to inappropriate operating conditions"
Instrument damage caused by electrostatic discharge
Electrostatic discharge (ESD) can damage the electronic components of the instrument
and the device under test (DUT). Electrostatic discharge is most likely to occur when
you connect or disconnect a DUT or test fixture to the instrument's test ports. To pre-
vent electrostatic discharge, use a wrist strap and cord and connect yourself to the
ground, or use a conductive floor mat and heel strap combination.
3.3
Bench Top Operation
If the analyzer is operated on a bench top, the surface must be flat. The instrument can
be used in horizontal or vertical position, standing on its feet, or with the support feet
on the bottom expanded.
Bench Top Operation