Basic operation
R&S
®
ZCxxx
29
Getting Started 1177.5156.02 ─ 05
Figure 6-1: Standard setup with two converters
4. Select "OK" to apply the converter configuration and close the dialog.
The R&S
ZNA now adjusts the frequencies and source power levels. The fre-
quencies correspond to the maximum frequency range of the configured con-
verters, the source power levels also take the cable and splitter losses into
account.
Furthermore, low phase noise mode is activated, and automatic level control
(ALC) is deactivated.
5. If you do not use the Rohde & Schwarz cables and accessories displayed in
, adjust the cable and splitter losses in the "Default Cable and Split-
ter Losses" dialog (System – [Setup] key > "Frequency Converter" tab >
"Default Cable and Splitter Losses ...").
Configuring the converter setup