Instrument Function
R&S
®
SMJ100A
433
Operating Manual 1403.7458.32 ─ 14
"Running"
The modulation signal is generated; a trigger was (internally or exter-
nally) initiated in triggered mode.
If "Armed_Auto" and "Armed_Retrigger" have been selected, genera-
tion of signals can be stopped with the "Arm" button. A new trigger
(internally with "Execute Trigger" or externally) causes a restart.
"Stopped"
The signal is not generated, and the instrument waits for a trigger
event (internal or external).
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger:RMODe?
Arm
Stops signal generation. This button appears only with "Running" signal generation in
the "Armed_Auto" and "Armed_Retrigger" trigger modes.
Signal generation can be restarted by a new trigger (internally with "Execute Trigger"
or externally).
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger:ARM:EXECute
Execute Trigger
Executes trigger manually. A manual trigger can be executed only when an internal
trigger source and a trigger mode other than "Auto" have been selected.
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger:EXECute
Trigger Source
Selects trigger source. This setting is effective only when a trigger mode other than
"Auto" has been selected.
"Internal"
The trigger event is executed by "Execute Trigger".
"External"
The trigger event is executed with the aid of the active edge of an
external trigger signal.
The trigger signal is supplied via the TRIGGER 1/2 connector.
The polarity, the trigger threshold and the input impedance of the trig-
ger input can be set in the "Global Trigger/Clock Settings" dialog.
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger:SOURce
Sync. Output to External Trigger
(enabled for Trigger Source External)
Enables/disables output of the signal synchronous to the external trigger event.
Baseband Signal - Baseband Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...