Instrument Function
R&S
®
SMJ100A
387
Operating Manual 1403.7458.32 ─ 14
"Next Segment
Seamless"
The segment selected under "Next Segment" is output.
This mode is only available if all segments have the same sample
rate.
●
In the case of "Internal Trigger = Auto", output starts at once and
the segment is generated continuously.
If the segment is changed in "Next Segment", the new segment is
output seamlessly after the output of the current segment is com-
plete.
●
In the case of "internal Trigger = Armed_Auto", output starts after
the first trigger event. The segment is then generated continu-
ously.
If the segment is changed in"Next Segment", signal output is
stopped and the new segment is not output until a trigger occurs.
●
In the case of "External Trigger = Auto", output starts at once and
the segment is generated continuously.
Each trigger event switches over to outputting the subsequent
segment once the output of the current segment has been com-
pleted.
If the segment is changed in "Next Segment", the new segment is
output seamlessly after the output of the current segment is com-
plete.
●
In the case of "External Trigger = Armed_Auto", output starts after
the first trigger event. The segment is then generated continu-
ously.
Each trigger event switches over to outputting the subsequent
segment once the output of the current segment has been com-
pleted.
If the segment is changed in "Next Segment", signal output is
stopped and the new segment is not output until a trigger occurs.
The remaining trigger modes ("Retrigger" and "Armed_Retrigger") are
not available.
"Sequencer"
The waveform files are processed according to the order and repeti-
tion cycles defined in the special sequencing list file (
*.wvs
)
assigned to the multi segment waveform file.
This “play list” is defined in the
Multi Segment Waveform Sequencing
dialog.
Note:
This feature is disabled in case the segments in the multi seg-
ment waveform file have different clock rates.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SMODe
Next Segment Source
Displays the next segment source, internal or external, as selected with the parameter
"Trigger Source".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:NEXT:SOURce?
Baseband Signal - Baseband Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...