Remote Control Commands
R&S
®
SMB100A
392
Operating Manual 1407.0806.32 ─ 21
Parameters:
<Mode>
SINGle | DOUBle | PTRain
SINGle
Enables single pulse generation.
DOUBle
Enables double pulse generation. The two pulses are generated
in one pulse period.
PTRain
A user-defined pulse train is generated The pulse train is defined
by value pairs of on and off times that can be entered in a pulse
train list.
*RST:
SINGle
Example:
PULM:MODE DOUB
enables double pulse generation.
Options:
R&S SMB-K23 (Pulse Generator), R&S SMB-K27 (Pulse Train)
Manual operation:
See
"Pulse Mode - Pulse Generator"
[:SOURce<hw>]:PULM:OUTPut:SYNC[:STATe]
<Sync>
Configures the signal at the SIGNAL VALID connector.
Parameters:
<Sync>
0 | 1 | OFF | ON
ON
Generates a single pulse at the beginning of a pulse sequence,
e.g. to synchronize pulse modulation.
OFF
Returns the validity of the RF signal at the output:
1
(high), while the signal settles.
0
(low), when it is stable (valid).
*RST:
OFF
Example:
PULM:OUTP:SYNC ON
uses the signal for synchronizing the pulse modulation.
Manual operation:
See
"Use SIGNAL VALID as Pulse Sync"
[:SOURce<hw>]:PULM:PERiod
<Period>
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.
Parameters:
<Period>
float
Range:
5 us | 20 ns to 100 s
Increment: 1us | 5 ns
*RST:
10 us
SOURce Subsystem