Remote Control Commands
R&S
®
SMB100A
288
Operating Manual 1407.0806.32 ─ 21
Example:
CAL:LEV:EXT:DATA FACT
selects the use of the data aquired at the factory for external
level correction.
Manual operation:
See
:CALibration<hw>:ROSCillator[:DATA]
<Data>
Sets the calibration value for the custom defined external adjustment.
Parameters:
<Data>
integer
Range:
0 to INT_MAX
*RST:
0
[:SOURce]:CALibration:STEReo:ANALog[:MEAS]?
The command starts all adjustments which affect the analog channels of the stereo
coder option.
Return values:
<Meas>
0 | 1
Example:
CAL:STER:ANAL?
starts the adjustments for analog channels of the stereo coder.
Response: 0
the adjustments have been performed successfully.
Usage:
Query only
Options:
R&S SMB-B5
Manual operation:
See
7.5 DIAGnostic Subsystem
The
DIAGnostic
system contains the commands used for instrument diagnosis and
servicing. SCPI does not define any DIAGnostic commands; the commands listed here
are all device-specific. All
DIAGnostic
commands are query commands which are
not influenced by
..............................................................................................288
:DIAGnostic<hw>:BGINfo:CATalog?
................................................................................ 289
..............................................................................................289
.......................................................................................... 290
:DIAGnostic<hw>:BGINfo?
[<Board>]
Checks the modules available in the instrument using the variant and revision state.
DIAGnostic Subsystem