Remote Control Commands
R&S
®
SMB100A
286
Operating Manual 1407.0806.32 ─ 21
Example:
SYST:FPR
All instrument settings (also the settings that are not currently
active) are reset to the factory values.
Usage:
Event
Manual operation:
See
7.4 CALibration Subsystem
The
CALibration
system contains the commands for adjustment. Adjustment is trig-
gered by the query commands. The response "0" indicates error-free adjustment, and
the response "1" means that an error occurred during adjustment.
:CALibration<hw>:ALL[:MEASure]?
................................................................................. 286
:CALibration<hw>:FMOFfset[:MEASure]?
.........................................................................286
:CALibration<hw>:FREQuency[:MEASure]?
......................................................................287
:CALibration<hw>:LEVel[:MEASure]?
...............................................................................287
:CALibration<hw>:LEVel:EXTern:DATA
............................................................................287
:CALibration<hw>:ROSCillator[:DATA]
............................................................................. 288
[:SOURce]:CALibration:STEReo:ANALog[:MEAS]?
........................................................... 288
:CALibration<hw>:ALL[:MEASure]?
[<Force>]
Starts all internal adjustments that do not require external measurement equipment.
Query parameters:
<Force>
string
Return values:
<Measure>
select
Example:
CAL:ALL:MEAS?
// Response "0"
// Adjustment has been performed successfully
Usage:
Query only
Manual operation:
See
:CALibration<hw>:FMOFfset[:MEASure]?
The command starts all adjustment for the FM/PhiM modulator.
Return values:
<Measure>
0 | 1
Example:
CAL:FMOF?
starts the adjustments for the FM/Phim modulator.
Response: "0"
the adjustments have been performed successfully
Usage:
Query only
CALibration Subsystem