Improving Level Performance
R&S
®
SMA100B
167
User Manual 1178.3834.02 ─ 03
"User"
The filter length is defined manually, with the parameter
As the filter length works as a multiplier for the time window, constant
filter length results in a constant measurement time.
Values 1 and 2N are allowed.
"Fixed Noise"
The averaging factor is selected so that the sensor's intrinsic noise (2
standard deviations) does not exceed the specified noise content.
Set the noise content value with the parameter
To avoid long settling times when the power is low, limit the averag-
ing factor with the parameter
.
Remote command:
:SENSe<ch>[:POWer]:FILTer:TYPE
Filter Length ← Sensor settings
Sets or indicates the filter length, depending on the selected filter mode.
●
"Filter > Auto" indicates the used filter length.
●
"Filter > User" enables you to adjust a manual optimal filter length.
●
"Filter > Fixed Noise" the filter length is blanked.
Remote command:
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER]
Noise/Signal Ratio ← Sensor settings
For "Filter > Fixed Noise", sets the noise content.
Remote command:
:SENSe<ch>[:POWer]:FILTer:NSRatio
Auto Once ← Sensor settings
Searches the optimum filter length for the current measurement conditions. The result
is indicated with the parameter
.
See also
"About the measuring principle, averaging filter, filter length, and achieving
Remote command:
:SENSe<ch>[:POWer]:FILTer:SONCe
Timeout ← Sensor settings
For "Filter > Fixed Noise", sets a time limit for the averaging process.
Remote command:
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
Default Aperture Time ← Sensor settings
The sensor default setting is sufficient. Disable this parameter to specify a user-defined
aperture time per sensor, if, for example, the readings vary.
To obtain stable readings, set the
exactly to one modulation period.
Remote command:
:SENSe<ch>[:POWer]:APERture:DEFault:STATe
Using Power Sensors