HP8360 emulation mode
1GP71_0E
90
Rohde & Schwarz
Status
Command / Function / Remark / Limitation
POWer:STEP[:INCRement] value
Set / Query the rf output level step size.
Value:
num
[DB] | | MAXimum | MINimum
0.0 dB
_
num
_
Pmax
Default:
10.0 dB
Remark:
MAXimum, MINimum are not supported.
POWer:STOP value
Set / Query the rf stop output level (in sweep mode).
Value:
num
[tbd] | UP | DOWN | MAXimum | MINimum
Pmin dBm
_
num
_
Pmax dBm
Default:
0.0 dBm
Remark:
UP, DOWN, MAXimum, MINimum are not supported.
Table 81: Power subsystem
•
PULM Subsystem
Status
Command / Function / Remark / Limitation
PULM:EXTernal:DELay value
PULM:EXTernal:POLarity value
PULM:INTernal:FREQuency value
PULM:INTernal:GATE value
PULM:INTernal:PERiod value
PULM:INTernal:TRIGger:SOURce value
PULM:INTernal:WIDTh value
PULM:SLEW value
PULM:SLEW:AUTO value
PULM:SOURce value
PULM:STATe value
Set / Query the (activation) state of the pulse modulation.
Value:
ON | OFF | 1 | 0
Default:
OFF