Remote Control Commands
R&S
®
SGT100A
316
User Manual 1176.8674.02 ─ 07
:CONNector:USER<ch>:CLOCk:IMPedance
<Impedance>
:CONNector:USER<ch>:TRIGger:IMPedance
<Impedance>
Selects the input impedance for the external trigger/clock inputs, when
is set to
TRIGger
or
CIN
/
COUT
.
Parameters:
<Impedance>
G50 | G10K
G10K
Provided only for backward compatibility with other R&S signal
generators.
The R&S
SGT accepts this values and maps it automatically to
G1K.
*RST:
G10K
Manual operation:
See
:CONNector:USER<ch>:CLOCk:SLOPe
<Slope>
:CONNector:USER<ch>:TRIGger:SLOPe
<Slope>
Sets the polarity of the active slope of an applied instrument trigger/clock.
Parameters:
<Slope>
NEGative | POSitive
*RST:
POSitive
Manual operation:
See
11.7 DIAGnostic Subsystem
..........................................................................................316
.......................................................................................316
:DIAGnostic:POINt:CATalog?
Queries the test points available in the instrument.
Usage:
Query only
:DIAGnostic[:MEASure]:POINt?
<Name>
Triggers voltage or temperature measurement at the specified test point and returns
the measured value.
Use the command
to retrieve a list of the available
test points.
For description of the test points, see the service manual.
DIAGnostic Subsystem