Signal Generator Settings
R&S
®
SGT100A
222
User Manual 1176.8674.02 ─ 07
2
= USB Adapter Micro-A to A
3
= Standard USB cable with USB type A and USB type B connectors
4
= USB type B connector
4a = External power supply unit, incl. power cable (supplied with the R&S NRP-Z5)
5
= Cable R&S NRP-ZK6 per sensor
Using USB hub
illustrates the connection as principle. For details, see the description
R&S
®
NRP
®
Series Power Sensors getting started.
Figure 7-26: Connecting power sensors to the R&S
SGT via USB hub
I, II =
G = Signal source (R&S
SGT)
1
= USB type Micro-A connector
2
= USB Adapter Micro-A to A
3
= Standard USB cable with USB type A and USB type B connectors
4
= USB type B connector
4a = External power supply unit and extra power cable
5
= Adapter cable R&S NRP-Z3 or R&S NRP-Z4 per sensor
Connection order
Always connect the equipment in the following order:
1. Connect the R&S NRP-Z5 sensor hub or the USB hub to the power supply and to
the R&S
SGT
2. Switch on the R&S
SGT
3. Connect/disconnect the R&S
NRP sensors
Detection and mapping
The R&S
SGT automatically detects a connected R&S
NRP power sensor and indi-
cates it in the "NRP Power Viewer" and "NRP Sensor Mapping" dialogs.
You can change the default mapping in the
About the measuring principle, averaging filter, filter length, and achieving sta-
ble results
A sensor measures the average signal power of the selected source continuously. The
measurement results are displayed the "NRP Power Viewer" dialog.
RF Settings