Signal Generator Settings
R&S
®
SGT100A
219
User Manual 1176.8674.02 ─ 07
Since the settings are interdependent, the affected parameters change accordingly if
you set a value.
To fill the table, press the "Execute" button.
To select the sensor and determine its parameters, refer to
Fill User Correction Data with Sensor
●
"Sensor"
Displays connected sensors for selection.
●
"List To Fill"
Indicates the used list.
●
"Include Zeroing"
Performs a zeroing procedure before acquiring the user correction data to improve
precision. Since during zeroing no signal may be applied to the sensor, RF is tem-
porarily switched off at the generator.
When unchecked, the zeroing procedure is skipped. However, the RF signal level
might be blanked shortly. This setting is recommended if blanking of RF is undesir-
able or the absence of power at the sensor cannot be guaranteed.
●
"Execute"
The "Execute" button is only enabled if a sensor is detected and the user correction list
contains at least one frequency value.
Remote command:
[:SOURce<hw>]:CORRection:ZERoing:STATe
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Used Instrument Settings For Measurement
Displays the settings relevant for the measurement.
"Modulation"
Indicates the modulation state
"Amplitude"
Shows the currently set level.
"Use SParameter"
Indicates whether SParameter correction is used.
"Attenuator Mode"
Displays the selected mode of the attenuator.
"Fixed range (PEP) In"
Shows the level range.
Remote command:
n.a.
7.7.7 NRP Sensor Mapping
The "NRP Sensor Mapping" lists all R&S
NRP sensors detected by the signal genera-
tor.
RF Settings