Signal Generator Settings
R&S
®
SGT100A
107
User Manual 1176.8674.02 ─ 07
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:PHASe
Generate Signal File
Generates a signal and saves it to a file.
The "File Select" window opens automatically and the signal can be stored as a wave-
form file.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:CREate:NAMed
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:CREate:NAMed
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:CIQ:CREate:NAMed
Generate Signal RAM
Generates a signal and uses it as output straight away.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:CREate
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:CREate
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:CIQ:CREate
7.1.2.6
Rect Test Signal
This dialog provides settings for configuration of a rectangular test signal. A rectangu-
lar test signal with a duty factor of 0.5 is created. Amplitude and offset are adjustable.
Both paths, I and Q, use the same signal.
Frequency
Enters the frequency of the test signal.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:FREQuency
Samples per Period
Enters the number of sample values required for the rectangular signal per period.
The maximum allowed value is determined by the maximum ARB clock rate (see data
sheet) and the selected frequency.
Baseband