Remote Control Commands
R&S
®
SGT100A
368
User Manual 1176.8674.02 ─ 07
Example:
BB:ARB:TSIG:RECT:SAMP 400
sets 400 samples per period
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:FREQuency
<Frequency>
Sets the frequency of the simple sinusoidal test signal. This signal is used as output via
the I channel. A sine wave of the same frequency but optionally phase-shifted is gener-
ated on the Q path (
ARB:TSIGnal:SINE:POFFset
).
Parameters:
<Frequency>
float
Range:
100 to depends on hardware
Increment: 0.01
*RST:
1000
Default unit: Hz
Example:
BB:ARB:TSIG:SINE:FREQ 100 kHz
sets a sine signal of 100 kHz.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:PHASe
<Phase>
The command sets the phase offset of the sine wave on the Q channel relative to the
sine wave on the I channel.
Parameters:
<Phase>
float
Range:
-180 to 180
Increment: 0.01
*RST:
90
Default unit: DEG
Example:
BB:ARB:TSIG:SINE:PHAS 90
sets a phase offset of 90 degrees.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:SAMPles
<Samples>
The command sets the sample rate for the sine signal in samples per period.
The maximum value is automatically restricted by reference to the set frequency and
has to fulfill the rule Frequency * Samples <= ARB clock rate.
Parameters:
<Samples>
integer
Range:
4 to 1000
*RST:
100
Default unit: 100 samples per period
SOURce Subsystem