Remote control commands
R&S
®
RTO6
2568
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n>
1..*
Selects the component. The DUT has suffix 1.
Parameters:
<InputPort1>
<InputPort2>
<OutputPort1>
<OutputPort2>
ONE | TWO | THRee | FOUR
Port number that is assigned to the input or output.
*RST:
ONE for input 1, TWO for output 1, THRee for
input
2, FOUR for output 4
Usage:
Asynchronous command
DEEMbedding<m>:COMPonent<n>:SPONe
<SParamFilePath>
DEEMbedding<m>:COMPonent<n>:SPTWo
<SParamFilePath>
Set the file name, file format and path of the S-parameter file, and load the S-parame-
ters.
See also:
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n>
1..*
Selects the component. The DUT has suffix 1.
Parameters:
<SParamFilePath>
String with path and file name with extension
s2p
(2-port com-
ponent) or
s4p
(4-port component).
Example:
See
Chapter 24.3.6.1, "Adding cables to the measurement
Usage:
Asynchronous command
24.24.4
Responses
DEEMbedding<m>:CADC?
Maximum attenuation or gain of the measurement setup before deembedding at DC.
This value is the expected basic attenuation or gain, which is corrected by deembed-
ding.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Deembedding (option R&S
RTO6-K121)