Remote control commands
R&S
®
RTO6
2480
User Manual 1801.6687.02 ─ 05
Table 24-19: Jitter measurements
Meas.
suffix
<MeasType>
parameter value
Meas. type
Description/Result
37
CCJitter
Cycle-cycle
jitter
Difference between the periods of two adjacent cycles. The
measurement is based on the period measurement.
38
NCJitter
N-cycle jitter
Difference between the time of two adjacent groups of N
cycles (periods) each.
39
CCWidth
Cycle-cycle
width
Difference between the pulse width of two adjacent cycles.
The measurement is based on the pulse width measurement.
40
CCDutycycle
Cycle-cycle
duty cycle
Difference between the duty cycle of two adjacent cycles.
The measurement is based on the duty cycle measurement.
41
TIE
Time interval
error
Time difference between the slope of the input signal and the
slope of a reference signal. The reference signal can be a
captured clock waveform, or a clock generated by clock data
recovery (CDR, software algorithm or hardware generation).
42
UINTerval
Unit interval
Period of the clock signal. If no clock signal is available, it is
recovered by CDR. The period is calculated as the time dif-
ference between two consecutive clock edges of the same
polarity.
43
DRATe
Data rate
Frequency of the clock signal. If no clock signal is available, it
is recovered by CDR. The measurement is based on the unit
interval measurement.
44
SKWDelay
Skew delay
Delay between the edges of two interdependent waveforms.
45
SKWPhase
Skew phase
Phase difference between the edges of two waveforms.
1 to 36;
46
Used for amplitude/time measurements (limit checks). The jitter category uses the same limit
checks as amplitude/time. See
Chapter 24.12.3, "Amplitude/time measurements"
..............................................................................2480
..................................................................................2481
...............................................................................2481
.............................................................................2482
MEASurement<m>:JITTer:SOURce<n>:TIESlope
........................................................... 2482
.................................................................................2483
MEASurement<m>:JITTer:SKWSlope
............................................................................ 2483
MEASurement<m>:JITTer:SKWRelation
.........................................................................2484
MEASurement<m>:JITTer:CCSLope
<JitterCCSlope>
Selects the slope at which the periods and thus the jitter is measured.
The command is available for the following measurements: cycle-cycle jitter, N-cycle
jitter, and cycle-cycle duty cycle (
are set to measurements CCJitter | NCJitter | CCDutycycle).
Jitter analysis and clock data recovery