Remote control commands
R&S
®
RTO6
2483
User Manual 1801.6687.02 ─ 05
The command is available for the following measurements: time-interval error, unit
interval and data rate (
are set to measurements TIE | UINTerval | DRATe.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
<n>
1..2
1 = data slope (only relevant for time interval error measure-
ments with explicit clock signal)
2 = clock slope
Parameters:
<JitterTIESlope>
POSitive | NEGative | EITHer
POSitive
The positive clock slope can be used, for example, for single
data rate (SDR) signals with bit start at the positive clock edge.
NEGative
The negative clock slope can be used, for example, for SDR sig-
nals with bit start at the negative clock edge.
EITHer
For clock edges, this option can be used for double data rate
(DDR) signals.
For data edges, it is the most common setting.
*RST:
EITHer
Usage:
Asynchronous command
MEASurement<m>:JITTer:DRUNit
<JittDatRateUnt>
Sets the unit for data rate measurements.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<JittDatRateUnt>
BIT_S | Hz
*RST:
BIT_S
Usage:
Asynchronous command
MEASurement<m>:JITTer:SKWSlope
<JitterSkewSlope>
Sets the edge of the first waveform from which the skew delay or phase is measured:
positive, negative or both.
Jitter analysis and clock data recovery