Remote control commands
R&S
®
RTO6
1295
User Manual 1801.6687.02 ─ 05
GPIB:TERMinator
<Terminator>
Specifies the symbol that is used as a terminator in GPIB communication.
Parameters:
<Terminator>
LFEoi | EOI
*RST:
EOI
Usage:
Asynchronous command
SYSTem:DISPlay:MESSage:STATe
<DispMessState>
Enables and disables the display of an additonal text in remote control.
.
Parameters:
<DispMessState>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
Firmware/software:
Version 2.70
SYSTem:DISPlay:MESSage[:TEXT]
<DisplayMessage>
Defines an additional text that is displayed during remote control operation.
To enable the text display, use
Parameters:
<DisplayMessage>
String that contains the text.
Usage:
Asynchronous command
Firmware/software:
Version 2.70
SYSTem:LANGuage
<Language>
Defines the remote control behavior of the instrument and sets the remote control com-
mand set.
Parameters:
<Language>
String value. Available values:
'SCPI': R&S
RTO6 remote command set is used.
"DPO7000" or 'TDS540': Compatible remote command set of
Tektronix oscilloscopes DPO7000 or TDS540 is used.
If one of these emulation modes is used, you can define alterna-
tive responses to the
IDN*?
and
OPT*?
commands on the
"Menu" > "Settings" > "System" > "Remote" > "SCPI Emulation"
tab.
Usage:
Asynchronous command
General remote settings