Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1138
User Manual 1801.6687.02 ─ 05
19.3.1.1
Power quality results
The results of " Power Quality" measurements are provided in two ways:
●
The diagram shows the graphical presentation of:
–
the voltage waveform
–
the current waveform
–
the power waveform that is the product of the current and voltage waveforms
●
The result box displays the numeric measurement results.
To measure and display the power quality, the instrument uses the following measure-
ments and waveforms:
●
"MG 6" to measure the voltage
●
"MG 7" to measure the current
●
"MG 8" to measure the power
●
"M2" "Math 2" to calculate the power
The used resources are listed in the "Details" tab. See also:
Voltage and current results
The voltage and current results are defined as follows:
Result
Description
RMS
Square root of the mean of the square of the current
or voltage averaged over N cycles
Peak
Highest measured magnitude value of the voltage or
current
Crest factor
Peak value / RMS value
Frequency
Frequency of the signal
Power results
The power in a system is described by several physical quantities: real power, reactive
power, complex power, and phase angle. In
, you can see how these quan-
tities are related if the voltage and the current are sinusoidal signals.