 R&S Scope Rider RTH Digital Oscilloscope 

Release Notes 1326.2645.00 - 5 

Version 

Function 

1.30.3.5 

Solved: Option K1, K2 threshold setting: It is necessary to first select the channel and only 
after that set the threshold. If the threshold is entered before selecting a new channel, the 
threshold value will not be applied. 

1.30.3.5 

Solved: For time bases between 500 s/DIV … 50 ms/DIV the instrument seems very slow as it 
only starts to display the trace after the acquisition has been fully completed and processed 
by firmware. 

1.20.3.4 

Options K1, K2 and K19: several problems solved with triggering, also in combinations of K1 
or K2 with K19 

1.20.3.4 

Solved: Option B1: Measurements on digital channels are not supported. 

1.20.3.4 

Solved: K15, History player cannot be stopped in mode automatic with 5000 segments 

1.20.3.4 

Solved: Option B1: In stop mode artefacts may be visible on digital channels. 

1.20.3.4 

Some issues with the autoset functionality were solved. 

1.20.3.4 

Solved: RTH1002, indication of statistics in frequency measurement (min, max, average over 
all frequencies) may not be correct in some cases 

1.20.3.4 

Solved: RTH1002/RTH1004 full selfalignment. In rare cases the AB alignment step fails. In 
this case repeat the full alignment. Typically the second try will pass. 

1.20.3.4 

Solved: Very rarely, if instrument is switched on and idle for a certain time: instrument reacts 
very slowly or not at all to keypad and touchscreen operation and needs to be rebooted. 

1.20.3.4 

Solved: Using Cursor with Math as input channel in combination with zoom is not supported. 
Combination can be selected, but leads to incorrect behavior of the instrument. 

1.10.2.10  Option B1, assignment of channel indicators corrected; user can now move digital channels 

on the screen 

1.10.2.10  Support added on RTH1002: temperature measurement using PT500 platinum sensor 

1.10.2.10  Improvement in trigger hysteresis, especially for 10:1 probe settings. 

1.10.2.10  Solved: Option K2, UART: the value of the parity bit is not correct 

1.10.2.10  Solved: Option K1, I2C bus triggers: trigger types address, data and address and data do not 

trigger reliably 

1.0.1.19 

Solved: Very sporadic incorrect trigger on the analog channels 

1.0.1.19 

Solved: The battery controller consumes very much energy from the battery when the RTH is 
switched off and disconnected from the mains. Therefore the battery runs empty very quickly. 

1.0.1.19 

Solved: Screenshot button: if pressed a second time before the pop-up window disappears 
the instrument will freeze. 

1.0.1.19 

Solved: Logger does not provide a backup of the log results yet if the RTH is switched off 
during logging. 

6

 

Known Issues 

The following table provides a more detailed overview about known issues. It also 
indicates the version in which the issue appeared for the first time. 

Version 

Function 

1.30.3.5 

None. 