Remote Commands Reference
R&S
®
RTC1000
234
User Manual 1335.7352.02 ─ 02
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Setting parameters:
<FileName>
String with path and file name
Usage:
Setting only
REFCurve<m>:LOAD
<FileName>
Loads the waveform data from the indicated reference file to the reference storage.
To load the instrument settings, use
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Setting parameters:
<FileName>
String with path and file name
Usage:
Setting only
REFCurve<m>:LOAD:STATe
Loads the instrument settings in addition to the reference waveform data. The wave-
form data must be loaded before the settings, see
.
The settings are only available if the file was stored to the internal storage
/INT/
REFERENCE
and never written to an external storage (USB stick).
Suffix:
<m>
.
1..4
Selects the reference waveform.
Usage:
Event
REFCurve<m>:HORizontal:SCALe
<Scale>
Changes the horizontal scale (timebase) of the reference waveform independent of the
channel waveform settings.
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Parameters:
<Scale>
Default unit: s/div