Error state
Meaning
Demagnetizing / automatic zero-
adjustment unavailable
This issue occurs when the DEMAG/
AUTO ZERO key is pressed under the
circumstances described in the table
below.
Demagnetization and automatic zero
adjustment cannot be performed. Per-
form demagnetization and zero adjust-
ment again after implementing the rem-
edy described below.
Circumstan-
ces
Remedy
When the
JAW
UNLOCKED
LED is lit up
Lock the sensor head.
(Press the opening lever
until the JAW
UNLOCKED indicator
turns off.)
When the
OVERLOAD
LED is flash-
ing
Remove the device from
the conductor being
measured.
When a cur-
rent of 0.5
A
(RMS) or
greater has
been detected