Remote Control Commands
R&S
®
NRX
360
User Manual 1178.5566.02 ─ 07
Manual operation:
See
[SENSe<Sensor>:]DM:STANdard
<standard>
Requires the sensor interface for R&S NRT (R&S NRX-B9).
Sets a communication standard for the modulation correction to reduce systematic
deviations occurring in power measurements.
Suffix:
<Sensor>
.
101
R&S NRT directional power sensor
Parameters:
<standard>
IS95 | WCDMa | DVBT | DAB
*RST:
IS95
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:RPLane
<plane>
Selects the power sensor port to which the measurement results are referred to.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<plane>
SOURce | LOAD
*RST:
SOURce
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:DMODulation[:VALue]
<modulation>
Sets a communication standard for the modulation correction to reduce systematic
deviations occurring in power measurements.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)