Remote Control Commands
R&S
®
NRX
355
User Manual 1178.5566.02 ─ 07
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
FLAT | NORMal | LTE | W3GPp
*RST:
FLAT
[SENSe<Sensor>:]BANDwidth[:RESolution][:VALue]
<value>
Effective for R&S
frequency selective power sensors.
Sets the resolution bandwidth.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
Range:
10.0 to 400.0e6
*RST:
25.0e6
Default unit: Hz
13.7.8
NRT Measurement Type
...............................................................................................355
....................................................................................... 359
................................................................................................362
13.7.8.1
NRT Mode Settings
Further information:
●
Chapter 8.5.1, "NRT Mode Settings"
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:MODE
................................ 355
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:PERiod
............................... 356
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:WIDTh
................................ 357
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:CCDF:THReshold
...........................357
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:DIRection
.......................................357
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:PEP:HOLD:TIME
............................358
INPut<Sensor>:PORT:SOURce:AUTO
............................................................................. 358
INPut<Sensor>:PORT:SOURce[:VALue]
...........................................................................359
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:MODE
<mode>
Defines how the average burst power is determined.
Configuring Sensors
www.allice.de
Allice Messtechnik GmbH