Remote Control Commands
R&S
®
NRX
328
User Manual 1178.5566.02 ─ 07
Usage:
Setting only
Manual operation:
See
"Frequency dependent offset active"
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet[:
MAGNitude]
<value>
Effective if
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:
ON
is set.
Considers the transmission loss in a cable that connects the desired measurement
point, set by
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: dB
Manual operation:
See
See
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:CORRection:STATe
<value>
Enables or disables gamma correction. See also
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
*RST:
OFF
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:PHASe
<value>
Effective if
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:
ON
is set.
Sets the phase angle of the complex reflection coefficient of the source.