Remote Control Commands
R&S
®
NRX
319
User Manual 1178.5566.02 ─ 07
Query parameters:
<tslot_width>
Default unit: s
<no_slots>
Default unit: s
<start_exclude>
Default unit: s
<end_exclude>
Default unit: s
<expected_value_or_source_list>
<resolution_or_source_list>
<source_list>
<expr>
Usage:
Query only
13.7
Configuring Sensors
...........................................................................................319
............................................................................................... 324
............................................................................................ 340
Frequency Selective Power Sensors
.................................................................... 353
....................................................................................... 355
13.7.1
Setting the Frequency
[SENSe<Sensor>:]FREQuency:FIXed
..............................................................................319
[SENSe<Sensor>:]FREQuency[:CW]
............................................................................... 319
[SENSe<Sensor>:]FREQuency:TRACk
............................................................................320
[SENSe<Sensor>:]FREQuency:FIXed
<frequency>
[SENSe<Sensor>:]FREQuency[:CW]
<frequency>
Sets the carrier frequency of the applied signal. This value is used for frequency-
response correction of the measurement result.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<frequency>
Range:
0.0 to 110.0e9
*RST:
1.0e9
Default unit: Hz
Manual operation:
See
Configuring Sensors