Remote Control Commands
R&S
®
NRX
223
User Manual 1178.5566.02 ─ 05
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
ON | OFF
*RST:
OFF
14.4.6.2
Combining Trace Commands
Parameter list
The following parameters are used.
●
<scope_size>
Mandatory. Number of test points on the time axis.
Corresponds to
●
<capture_time>
Mandatory. Period within which measured data are captured in the trace measure-
ments.
Corresponds to
●
<source_list>
Mandatory. Primary and secondary sensor. The number for each sensor is prece-
ded by the character
@
. The entire expression is enclosed in parentheses.
Example: (
@
3),(
@
2)
Sensor C is the primary sensor, and sensor
B is the secondary sensor.
CONFigure<Measurement>:XTIMe[:POWer]?
<scope_size>, <capture_time>,
<source_list>
FETCh<Measurement>:XTIMe[:POWer]?
<scope_size>, <capture_time>,
<source_list>
READ<Measurement>:XTIMe[:POWer]?
<scope_size>, <capture_time>,
<source_list>
MEASure<Measurement>:XTIMe[:POWer]?
<scope_size>, <capture_time>,
<source_list>
Used to measure power over time.
The used parameters are described in
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Query parameters:
<scope_size>
<expr>
<capture_time>
Default unit: s
<source_list>
<expr>
Usage:
Query only
Measurement Settings and Results