Remote Control Commands
R&S
®
NRX
207
User Manual 1178.5566.02 ─ 05
TRIGger<Measurement>[:CHANnel<Channel>]:SLOPe
.....................................................213
........................................................................................214
TRIGger<Measurement>[:CHANnel<Channel>]:SOURce
...................................................214
TRIGger<Measurement>[:CHANnel<Channel>]:SYNChronize:PORT
..................................214
TRIGger<undef>:ALL:SYNChronize[:STATe]
.....................................................................215
TRIGger<Measurement>[:CHANnel<Channel>]:SYNChronize[:STATe]
................................215
TRIGger<undef>:ALL[:IMMediate]
................................................................................... 215
TRIGger<Measurement>[:IMMediate]
.............................................................................. 215
TRIGger<undef>:ALL:ATRigger[:STATe]
<stat>
TRIGger<Measurement>[:CHANnel<Channel>]:ATRigger[:STATe]
<stat>
Controls the automatic trigger function. If enabled, an artificial trigger is generated if the
delay time has elapsed after the measurement start and no trigger event has occurred.
The delay time is set using
TRIGger<Measurement>[:CHANnel<Channel>]:
.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<stat>
ON | OFF
*RST:
OFF
TRIGger<undef>:ALL:COUNt
<count>
TRIGger<Measurement>[:CHANnel<Channel>]:COUNt
<count>
Sets the number of measurement cycles to be performed when the measurement is
started using
INITiate<Measurement>[:IMMediate]
.
This number equals the number of results that can be obtained from the sensor after a
single measurement. As long as the defined number of measurements is not executed,
the sensor automatically initiates another measurement internally when the current
result is available.
This command is particularly useful in conjunction with buffered measurements. For
example, to fill a buffer with a predefined size with measurements that have been trig-
gered externally or by
without having to start the measurement multiple times.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Measurement Settings and Results