Instrument functions
R&S
®
NGU
47
User Manual 1179.2531.02 ─ 03
Figure 6-3: Constant resistance dialog
6.2.2
Fast transient response
With fast transient response, the R&S
NGU is able to quickly stabilize the output volt-
age upon a step change in the load current.
Load transient recovery time can be switched between 30
μ
s ("Fast Transient
Response" = "ON") or 100
μ
s ("Fast Transient Response" = "OFF") for the output volt-
age to recover and stay within 20 mV of the nominal output voltage follows by a step
change in the load current.
Figure 6-4: Transient response graph
1. Press [Settings] key.
The R&S
NGU displays the device/channel menu window.
2. Select the channel tab to apply fast transient response.
The R&S
NGU displays the channel menu.
Activating the channel output